Description
The
hThyroglobulin [125I] IRMA system provides a direct
quantitative in vitro determination of human thyroglobulin (hTg) in
human serum. hTg can be assayed in the range 0-250 ng/ml using 100 µl
serum sample. Each kit contains materials sufficient for 100 assay tubes
permitting the construction of one standard curve and the assay of 40
unknowns in duplicate (20 sample with recovery test in duplicate).
Introduction
Thyroglobulin is a iodoglycoprotein consisting of heterogeneous
molecules, the composition of which is in part depending on the degree
of iodination. The prevailing molecular form is 660 kDa (dimeric form,
the two subunits, linked by noncovalent bounds), but both larger and
smaller molecular forms exist in the thyroid gland. Tg is the site of
synthesis and storage of thyroid hormone. produced by the thyroid gland.
Tg is synthesized and stored in thyroid follicles and some of the
nonenzimatically digested protein is released into the circulation upon
stimulation with thyrotropin (TSH) together with thyroxine (T4)
and triiodothyronine (T3).
The
determination of Tg by immunoassay methods plays a crucial role in the
diagnosis of thyroid disorders, such as Grave’s multinodular goiter,
benign thyroid adenoma, thyroiditis (acute phase), and differentiated
carcinoma, and is a useful tool in screening population at risk for
thyroid carcinoma after previous irradiation.
The
sensitivity of the present hTg IRMA system makes it suitable for the
measurement of subnormal hTg levels, which is an early and reliable
marker of tumor recurrence.
Principle of method
The technology
uses two high affinity monoclonal antibodies in an immunoradiometric
assay (IRMA) system.
The 125I
labelled signal-antibody binds to an epitope of the Tg molecule
spatially different from that recognised by the biotin-capture-antibody.
The two antibodies react simultaneously with the antigen present in
standards or samples, which leads to the formation of a capture antibody
- antigen - signal antibody complex, also referred to as a “sandwich”.
During an
overnight incubation period the immuno-complex is immobilized to the
reactive surface of streptavidin coated test tubes. Reaction mixture is
then discarded, test tubes washed exhaustively, and the radioactivity is
measured in a gamma counter.
The
concentration of antigen is directly proportional to the radioactivity
measured in test tubes. By constructing a calibration curve plotting
binding values against a series of calibrators containing known amount
of hTg, the unknown concentration of hTg in patient samples can
determined.
Contents of the kit
|
1
bottle |
TRACER, ready to use.
21 ml, containing about 980 kBq 125I-anti-hTg and
capture anti-hTg in buffer with red dye and 0.1% NaN3. |
|
6
vials |
STANDARD, ready to use.
1 ml per vial, containing 0.3 (S1), 1.0 (S2),
4.0 (S3), 20 (S4) 100 (S5) and
250 (S6) ng/ml hTg (BCR CRM457) in human serum with
0.1% NaN3 |
|
2
vials |
CONTROL SERUM
1.0 ml, containing 0.1% NaN3
The concentration of the control sera is specified in the
quality certificate enclosed. |
|
1 vial |
SERUM DILUENT
5.0 ml, containing 0.1% NaN3 |
|
2
boxes |
COATED TUBE, ready to use.
2 x 50 reactive test tubes, 12x75 mm, packed in plastic boxes. |
|
1
bottle |
WASH BUFFER CONCENTRATE.
20 ml, containing 0.1% NaN3. See Preparation of
reagents |
|
1 pc |
Quality certificate |
|
1 pc |
Pack
leaflet |
Materials, tools and equipment required
Test
tube rack
Precision pipettes with disposable tips (100, 200 and 2000 µl)
Distilled water
Vortex mixer
Shaker
Plastic foil
Adsorbent tissue
Gamma counter
Recommended tools and equipment
Repeating pipettes (e.g., Eppendorf)
Dispenser with 1-L reservoir (instead of the 2 ml pipette)
Specimen collection and storage
Serum
samples can be prepared according to common procedures used routinely in
clinical laboratory practice. Samples can be stored at 2-8 °C if the
assay is carried out within 24 hours, otherwise aliquots should be
prepared and stored deep frozen (-20 °C). Frozen samples should be
thawed and thoroughly mixed before assaying. Repeated freezing and
thawing should be avoided. Do not use lipemic, hemolyzed or turbid
specimens.
Samples
with a thyroglobulin concentration higher than that of the most
concentrated standard should be diluted and reassayed.
Preparation of reagents, storage
Add the
wash buffer concentrate (20 ml) to 700 ml distilled water to obtain 720
ml wash solution. Upon dilution store at 2-8 °C until expiry date.
Store
the rest of reagents between 2-8 °C after opening. At this temperature
each reagent is stable until expiry date. The actual expiry date is
given on the package label and in the quality certificate.
CAUTION! Equilibrate all reagents and serum samples to room temperature.
Mix all reagents and samples thoroughly before use. Avoid excessive
foaming.
Recovery test
Anti-Tg
antibodies or unspecific effects in a patient’s serum can interfere with
serum thyroglobulin measurement, which leads to underestimation of the
Tg concentration in IRMA system. Interference can be detected by using
recovery test. The recovery test should be carried out as described in
the Assay procedure.
The
concentration of the recovery sample (approximately 500ng Tg/ml) should
be checked with serum diluent (recovery reference tubes (DR)).
Recovery (in %) in the serum sample:
|
ng Tg/ml Rx – ng Tg/ml Sx |
x100 = % recovery |
|
—————————— |
|
ng
Tg/ml DR |
Recoveries between 70% and 130% are considered valid. Levels of <70% or
>130% are due to interference and the Tg level of the relevant original
sample is considered invalid.
Assay procedure
(For a
quick guide, refer to Table 1.)
|
1 |
Equilibrate reagents and samples to room temperature before use. |
|
2 |
Label
coated tubes in duplicate for each standard (S1-S6), control
serum (CI, CII), serum diluent (D) as zero calibrator, recovery
reference (DR), serum samples (Sx), and recovery sample (Rx).
|
|
3 |
Homogenize all reagents and samples by gentle mixing.
Avoid foaming. |
|
4 |
Pipette 10 µl recovery sample into the recovery reference
tubes (DR) and into the sample recovery tubes (R).
|
|
5 |
Pipette 100 µl of standards into the standard tubes
(S1-S6), 100 µl control into control tubes (CI, CII), 100µl
sample into sample (S) and recovery tubes (R) and 100 µl serum
diluent into the recovery reference tubes (DR) and serum diluent
tubes (D) as zero calibrator. Use rack to hold the tubes. Do not
touch or scratch the inner bottom of the tubes with pipette tip.
|
|
6 |
Pipette 200 µl of tracer into each tube. |
|
7 |
Gently
vortex all tubes. Seal all tubes with a plastic foil.
|
|
8 |
Incubate tubes for 15-24 hours at 25 °C. |
|
9 |
Add 2.0 ml diluted wash buffer to each tube and decant
the supernatant from all tubes by the inversion of the rack. In
the upside down position place the rack on an absorbent paper
for 2 minutes. |
|
10 |
Return the tube rack to an upright position, and repeat
Step 9 two times more |
|
11 |
Count each tube for at least 60 seconds in a gamma
counter. |
|
12 |
Calculate the Tg concentrations of the samples as
described in Calculation of results or use special
software. |
Note
for the washing step: Decantation is the most critical step of
the assay procedure. Pay a special attention not to contaminate the
outer surface of tubes, when turning the test tube-rack upside down.
Even a small contamination may introduce a high, unidentified background
resulting in a substantial over-estimation of concentration. The error
associated may become particularly high in the low range of
concentration, which is of vital importance for the reliable
determination of subnormal Tg-values. On the same reason, regular
checking of the instrument background is indispensable. This is
particularly important, when multi-channel counters are used. Make
ensure that background values and variation between individual channels
are within the range of acceptance as specified in counter's service
book.
Table
1. Assay Protocol, Pipetting Guide (all volumes in microliters)
|
Tube
Reagents |
Total
(T) |
Serum diluent
(D) |
Standard
(S1-S6) |
Sample (Sx) |
Recovery tubes (Rx) |
Recovery reference tubes (DR) |
Control serum
(CI-CII) |
|
Standard |
|
|
100 |
|
|
|
|
|
Sample |
|
|
|
100 |
100 |
|
|
|
Control serum |
|
|
|
|
|
|
100 |
|
Recovery sample |
|
|
|
|
10 |
10 |
|
|
Serum
diluent |
|
100 |
|
|
|
100 |
|
|
Tracer |
200 |
200 |
200 |
200 |
200 |
200 |
200 |
|
Incubate tubes for 15-24 hours at 25 oC |
|
Wash
buffer |
2000 |
2000 |
2000 |
2000 |
2000 |
2000 |
|
Decant
the fluid and blot on filter paper |
|
Repeat the washing step twice |
|
Counting radioactivity (60 sec/tube) |
|
Calculation |
Calculation of results
The
calculation is illustrated using typical data. The assay data collected
should be similar to those shown in Table 2.
Calculate the
average CPM for each pair of assay tubes. Calculate the normalized
percent binding for each standard, control and sample respectively by
using the following equation:
| |
S1-6 [C, Sx, Rx] (cpm) |
|
|
B / T % = |
————————— |
x 100 |
| |
T (cpm) |
|
For
simplicity, these values are uncorrected for non-specific binding (NSB).
This is enabled by low NSB being less than 3% of total count.
Using
semi-logarithmic graph paper plot B/T (%) for each standard versus the
corresponding concentration of Tg.
Determine the Tg concentration of the unknown samples by interpolation
from the standard curve. Do not extrapolate values beyond the standard
curve range.
Out of
fitting programs applied for computerized data processing logit-log, or
spline fittings can be used. Automated data processing systems are also
available.
Figure 1.
Typical standard curve
(Do not use to calculate sample values)
Table 2.
Typical assay data
| |
Tg
ng/ml |
cpm
1 |
cpm
2 |
Cpm
mean |
Tg
ng/ml |
|
Total |
- |
393128 |
394123 |
393626 |
- |
|
D
(NSB) |
0 |
167 |
178 |
171 |
|
|
S0.3 |
0.3 |
552 |
559 |
556 |
- |
|
S1.0 |
1.0 |
1448 |
1476 |
1462 |
- |
|
S4.0 |
4.0 |
5517 |
5500 |
5509 |
- |
|
S20 |
20 |
24756 |
24712 |
24734 |
- |
|
S100 |
100 |
90031 |
90207 |
90119 |
- |
|
S250 |
250 |
158821 |
162358 |
160590 |
- |
|
CI |
- |
2458 |
2502 |
2480 |
1.99 |
|
CII |
- |
70250 |
70511 |
70380 |
69.8 |
Characterization of the assay
Typical assay parameters
|
NSB / T |
|
< 0.075% |
|
Bmax / B0 |
|
> 500 |
Sensitivity
The
analytical sensitivity or minimum detectable limit is calculated by the
interpolation of the mean counts of zero standard plus 2 standard
deviation from the standard curve. Determination was carried out using
20 replicates of zero standard response.
The
value of analytical sensitivity is 0.022 ng/ml measured using fresh
tracer. The declared analytical sensitivity is better then 0.1 ng/ml in
any time before expiry.
Hook
effect
There is no
high dose hook effect up to an hTg concentration 20000 ng/ml.
Linearity – dilution test
Three
individual human serum samples were diluted with the zero standard of
the kit. The diluted samples were measured according to kit protocol.
|
sample
No. |
dilution factor |
expected ng/ml |
observed ng/ml |
recovery % |
|
1 |
|
68.5 |
68.5 |
|
|
1 |
2 |
34.0 |
34.3 |
101.0 |
|
1 |
4 |
16.7 |
17.0 |
101.6 |
|
1 |
8 |
8.2 |
8.3 |
101.0 |
|
1 |
16 |
4.1 |
4.2 |
104.2 |
|
2 |
|
93.6 |
93.6 |
|
|
2 |
2 |
46.4 |
45.2 |
97.5 |
|
2 |
4 |
22.9 |
22.0 |
96.3 |
|
2 |
8 |
11.3 |
10.7 |
95.2 |
|
2 |
16 |
5.6 |
5.3 |
95.2 |
|
3 |
|
67.5 |
67.5 |
|
|
3 |
2 |
33.6 |
31.8 |
94.7 |
|
3 |
4 |
16.6 |
15.4 |
93.0 |
|
3 |
8 |
8.2 |
7.5 |
92.3 |
|
3 |
16 |
4.0 |
3.7 |
93.0 |
Recovery – addition test
49 individual
human serum samples were spiked with known concentration serum based
stock solution made from BCR certification reference preparation 457 in
different amount. The results are summarised below.
|
sample
base concentration (ng/ml) |
concentration with Recovery sample (ng/ml) |
Recovery% |
|
Sample
1 |
6.9 |
60.6 |
101 |
|
Sample
2 |
15.5 |
69.9 |
103 |
|
Sample
3 |
99.5 |
152.6 |
100 |
|
Sample
4 |
2.3 |
58.8 |
107 |
|
Sample
5 |
10.4 |
63.5 |
100 |
|
Sample
6 |
57.3 |
109.0 |
97 |
|
Sample
7 |
6.5 |
62.0 |
105 |
|
Sample
8 |
7.0 |
61.0 |
102 |
|
Sample
9 |
2.5 |
52.8 |
95 |
|
Sample
10 |
106.0 |
159.4 |
101 |
|
Sample
11 |
32.4 |
83.8 |
97 |
|
Sample
12 |
7.4 |
60.8 |
101 |
|
Sample
13 |
41.0 |
88.2 |
89 |
|
Sample
14 |
8.8 |
63.9 |
104 |
|
Sample
15 |
0.9 |
56.9 |
106 |
|
Sample
16 |
12.9 |
61.0 |
91 |
|
Sample
17 |
0.3 |
51.1 |
96 |
|
Sample
18 |
6.2 |
61.3 |
104 |
|
Sample
19 |
3.9 |
58.7 |
103 |
|
Sample
20 |
2.0 |
51.6 |
94 |
|
Sample
21 |
3.4 |
52.2 |
92 |
|
Sample
22 |
3.3 |
55.7 |
99 |
|
Sample
23 |
43.9 |
91.8 |
90 |
|
Sample
24 |
10.1 |
62.7 |
105 |
|
Sample
25 |
8.6 |
62.9 |
109 |
|
Sample
26 |
0.6 |
36.8 |
72 |
|
Sample
27 |
0.3 |
54.8 |
109 |
|
Sample
28 |
11.4 |
52.3 |
82 |
|
Sample
29 |
10.3 |
63.8 |
107 |
|
Sample
30 |
7.9 |
64.5 |
113 |
|
Sample
31 |
0.6 |
57.2 |
113 |
|
Sample
32 |
33.3 |
76.3 |
86 |
|
Sample
33 |
5.0 |
64.8 |
120 |
|
Sample
34 |
11.5 |
68.3 |
114 |
|
Sample
35 |
22.9 |
71.0 |
96 |
|
Sample
36 |
2.3 |
50.4 |
96 |
|
Sample
37 |
7.9 |
53.9 |
92 |
|
Sample
38 |
0.5 |
39.5 |
78 |
|
Sample
39 |
5.4 |
48.2 |
86 |
|
Sample
40 |
17.2 |
73.7 |
113 |
|
Sample
41 |
190.8 |
235.4 |
89 |
|
Sample
42 |
77.5 |
132.1 |
109 |
|
Sample
43 |
5.1 |
61.5 |
113 |
|
Sample
44 |
23.1 |
66.2 |
107 |
|
Sample
45 |
76.4 |
119.4 |
107 |
|
Sample
46 |
4.7 |
47.2 |
106 |
|
Sample
47 |
154.2 |
199.1 |
112 |
|
Sample
48 |
2.3 |
44.2 |
104 |
Intra-interassay
|
intraassay |
interassay |
|
mean
(ng/ml) |
CV % |
mean
(ng/ml) |
CV % |
|
11.4 |
1.8 |
11.5 |
2.2 |
|
118.7 |
1.9 |
117.8 |
2.7 |
|
85.4 |
2.2 |
86.2 |
1.9 |
|
5.9 |
2.6 |
6.0 |
1.7 |
|
4.9 |
2.0 |
5.1 |
2.0 |
|
1.3 |
2.7 |
1.3 |
3.0 |
|
0.7 |
5.9 |
0.7 |
6.3 |
Expected Values
Expected
normal value range is 2 ng/ml - 70 ng/ml.
It is
recommended that each laboratory determine a reference range for healthy
persons for its own patient population, since this may vary in different
laboratories or regions.
Procedural notes
1)
Source of error! Reactive test tubes packed in plastic
boxes are not marked individually. Care should be taken of not mixing
them with common test tubes. To minimize this risk, never take more
tubes than needed out of plastic box, and put those left after work back
to the box. It is recommended to label assay tubes by a marker pen.
2)
Source of error! To ensure the efficient rotation,
tubes should be firmed tightly inside the test tube rack. Never use a
rack type with open hole. An uneven or incomplete shaking may result in
a poor assay performance.
3)
Addition of wash buffer. For the addition of wash
buffer the use of a common laboratory dispenser equipped with a 1-L
glass bottle, and a flexible outlet tubing end is recommended. In lack
of this tool a large volume syringe attached to a repeating pipette can
be used.
Additional information
Components from various lots or from kits of different manufacturers
should not be mixed or interchanged.
Precaution
Radioactivity
This product
contains radioactive material. It is the responsibility of the user to
ensure that local regulations or code of practice related to the
handling of radioactive materials are satisfied.
Biohazard
Human
blood products used in the kit have been obtained from healthy human
donors. They were tested individually by using approved methods (EIA,
enzyme immunoassay), and were found to be negative, for the presence of
both Human Immunodeficiency Virus antibody (Anti-HIV-1) and Hepatitis B
surface Antigen (HBsAg).
Care
should always be taken when handling human specimens to be tested with
diagnostic kits. Even if the subject has been tested, no method can
offer complete assurance that Hepatitis B Virus, Human Immunodeficiency
Virus (HIV-1), or other infectious agents are absent. Human blood
samples should therefore be handled as potentially infectious
materials.
Chemical hazard
Components contain sodium azide as an antimicrobial agent. Dispose of
waste by flushing with copious amount of water to avoid build-up of
explosive metallic azides in copper and lead plumbing. The total azide
present in each pack is 74 mg. |